Understanding Hemodialysis Water Quality Standards
Water quality is the most critical and most frequently overlooked factor in dialysis safety. Each patient session exposes them to 120–150 litres of dialysate that must meet strict microbiological and chemical purity standards.
Key parameters to monitor in your RO system
Below 10 µS/cm
Readings above this at point of use indicate membrane degradation or bypass contamination.
Below 100 CFU/mL
AAMI standard. EUDRACT stricter guideline requires below 50 CFU/mL for dialysis water.
Below 0.25 EU/mL
Standard dialysis limit. Online HDF requires below 0.03 EU/mL — ultrapure grade.
5.0 – 7.0
Required pH range for RO permeate in dialysis applications. Outside this range accelerates membrane wear.
Common causes of RO membrane fouling
- Inadequate pre-treatment — hardness, chlorine, or iron in feed water
- Insufficient disinfection frequency
- Delayed membrane replacement beyond rated lifespan
- Bypassed or non-functional UV lamp